From Model to Product: The Implementation Gap

Why most published AI models never reach clinical practice. The chasm between research performance and real-world deployment, and what it takes to cross it.


There are tens of thousands of published papers describing AI models for medicine, and a much smaller number of AI tools that a clinician somewhere uses on a normal working day. The FDA's public list of AI-enabled medical devices passed a thousand entries at the end of 2024 and had about 1,600 by late 2026, which sounds like a lot until you compare it with the literature, and being on the list only means a product may be sold. Plenty of cleared products have very few users.

The distance between a model that works in a study and a tool that is used in care is usually called the implementation gap. 1.6 What AI Cannot Do covered the scientific reasons a model disappoints when it leaves the lab. This article is about the rest: the engineering, organizational and financial work that sits between a validated model and routine use, most of which has nothing to do with machine learning.

How big is the gap?

In 2020 Myura Nagendran and colleagues reviewed the studies that had compared deep learning with clinicians on medical imaging tasks. Of 81 non-randomized studies, nine were prospective and six had been tested in a real clinical setting. They found ten randomized trials registered, of which two had been published. The typical study was retrospective, run on a dataset the authors already had, and that is roughly where most models stop.

We shouldn't be too hard on the authors. A retrospective accuracy study is a reasonable first step, and it is the step a research grant pays for. The trouble is that the later steps are bigger than the first one, and each needs skills and money that a research lab mostly doesn't have.

The model is a small part of the system

In 2015 a group of Google engineers led by D. Sculley published a paper called "Hidden Technical Debt in Machine Learning Systems." Its best-known figure shows a production machine learning system as a set of boxes, with the model code as one small box in the middle, surrounded by much larger ones for data collection, data verification, serving infrastructure, configuration and monitoring.

That figure was drawn for ad-ranking systems, and it applies to clinical AI with a few more boxes added. To run a model in a hospital, somebody has to build or obtain all of this:

  • A way to receive the right studies from the hospital's systems automatically, and only the right ones.
  • The same preprocessing that was used in training, applied to whatever the scanners produce today.
  • Checks that reject inputs the model wasn't built for: the wrong body part, the wrong sequence, a pediatric patient for an adult model.
  • Somewhere to run the model, on hardware the hospital will approve, with a response time that fits the workflow.
  • A way to get the result back in front of the right person, in the software they already use.
  • Logging, version control, security, and a plan for when it breaks at 2 a.m.

A research model has none of these. It has a script that reads a folder of files that somebody cleaned by hand.

Why models stall

The evidence is the wrong kind

A retrospective AUC answers the question "can the model tell these cases apart?" The people who decide on adoption are asking different questions. Does it work on our patients and our scanners? Do our clinicians do better with it than without it? Does anything improve for the patient? Those need external validation, reader studies and prospective studies, which are covered in 5.1 Technical Validation vs. Clinical Validation.

It doesn't fit the work

A tool that requires a radiologist to open a second application, log in and upload a study will not be used, however accurate it is. The result has to arrive inside the existing workflow, at the moment a decision is being made, addressed to someone who can act on it. Working out where that moment is takes actual study of the workflow (see 8.2 Clinical Workflow Analysis), and connecting to the hospital's systems takes integration work with standards most researchers have never had to touch (see 8.3 Interoperability).

When a Google Health team took their diabetic retinopathy model into clinics in Thailand, the model was the part that worked. The problems they reported in 2020 were poor lighting in the screening rooms, slow internet connections that held up each patient, and a referral process that didn't suit patients who had travelled a long way. None of that is visible from a test set.

It is a medical device

Software that informs diagnosis or treatment is generally regulated, and you can't deploy it for clinical use because the paper was accepted. Clearance needs a quality management system, design documentation, risk management and a validation package (section 6 of this knowledge base, starting at 6.1 When Does Your AI Algorithm Become a Medical Device?). The awkward part for a research group is that much of this documentation is supposed to be created during development. Research code written with no version control and no record of which data trained which model usually can't be documented after the fact, and gets rebuilt.

Nobody is paying for it

This is the reason that gets the least attention in papers and probably stops the most projects. A hospital that adopts an AI tool pays for the licence, the integration, the IT security review, the training and the ongoing support. In most health systems there is no payment for using AI as such, so the tool has to pay for itself through time saved, extra throughput or avoided harm, and somebody has to make that case with numbers.

There are a few exceptions in the United States. In 2020 Medicare granted a new technology add-on payment for Viz.ai's software for detecting large-vessel occlusion strokes, and since 2021 there has been a billing code for retinal imaging with automated analysis at the point of care, the use pioneered by IDx-DR, which in 2018 became the first device the FDA authorized to give a screening result without a clinician also interpreting the image. They are notable mostly for being rare.

Savings are also tricky because they often land in a different budget from the costs. A tool that reduces unnecessary biopsies saves the payer money and reduces the hospital's revenue.

People don't trust it, or trust it too much

Clinicians who weren't involved in building a tool, don't know how it was validated and have seen it be confidently wrong once will ignore it. Clinicians who trust it completely are a different problem. Both are covered in 8.4 Human-AI Interaction and User Interface Design and 8.5 Clinician Trust, Adoption, and Change Management.

It needs looking after

A deployed model is never finished. Scanners get replaced, protocols change, the patient mix moves, and performance moves with them. Somebody has to watch for that and have the authority to act (see 8.6 Monitoring AI Performance in Production). A research lab whose postdoc has moved on is not well placed to do it.

We have done this before

Computer-aided detection for mammography was approved by the FDA in 1998, Medicare began paying extra for it in 2002, and by 2008 it was used on about three quarters of the screening mammograms Medicare paid for. In 2015 Constance Lehman and colleagues analysed more than 600,000 mammograms read with and without it and found no improvement in diagnostic accuracy. Sensitivity was no better, and for radiologists who read both ways it was a little worse with CAD.

So getting across the gap is not the same thing as benefit. That generation of software was adopted because it was reimbursed, on the strength of early studies that didn't reflect real practice. The lesson we take from it is that the evidence questions and the implementation questions have to be answered together. Adoption without evidence of benefit is not a success.

What it takes

The stages below are roughly in order, though in practice they overlap and loop back.

  1. A clinical problem that someone owns. A named clinical lead who wants the problem solved, a defined decision the tool will inform, and a measure of what "better" would look like.
  2. A model validated on data from outside the development site, with subgroup results.
  3. A fixed, documented version of the model and its preprocessing, built under design controls if clinical use is the goal.
  4. Integration with the systems where the work happens.
  5. A silent period. The tool runs on live data with its output hidden from clinicians, so you can check performance and reliability on local patients without affecting care.
  6. A staged rollout with training, a channel for feedback and a baseline measurement to compare against.
  7. A prospective evaluation of the effect on decisions, workload or outcomes.
  8. Monitoring and maintenance, with a named owner and a budget.

Almost none of these steps need a machine learning researcher. They need software engineers, a regulatory and quality person, the hospital's IT and PACS teams, clinical leads, and somebody managing the whole thing. This is why most models that reach practice get there through a company, either a spin-out or a licensing deal, and why a research group that wants its work used should talk to its technology transfer office earlier than feels natural.

Things you can do while it is still a research project

You don't have to become a company to improve your model's chances. A few habits during the research phase make the later stages possible.

Choose the problem with a clinician who will use the answer. A model built because the dataset was available tends to answer a question nobody was asking.

Write down the intended use early: which patients, which images, which user, what decision. It shapes the dataset, the validation and eventually the regulatory route (see 1.5 Matching the Problem to the Approach).

Keep records as if someone will audit them. Which data trained which model version, with which code and settings. This costs little while you are doing the work and is very hard to reconstruct later (see 7.5 Transparency and Reproducibility).

Train on the data as it comes out of the clinical systems. A model that only works on hand-curated research files will need redoing.

Check the permissions. Ethics approval and data agreements written for research may not cover commercial development (see 2.7 De-Identification, Privacy, and Regulatory Compliance for Research Data).

And test it somewhere else before you believe it.

Key takeaways

  • Most published models never reach routine clinical use. The typical study is retrospective and single-site, and the work that comes after it is larger than the work before.
  • The model is a small component of a deployed system. Data intake, input checks, integration, infrastructure, logging and support make up most of it.
  • Models stall for reasons of evidence, workflow fit, regulation, money, trust and maintenance, and usually several at once.
  • Reimbursement decides a lot. A tool has to pay for itself or be paid for, and the savings often fall in someone else's budget.
  • Mammography CAD was widely adopted and later shown not to improve accuracy. Adoption and benefit have to be shown separately.
  • A silent period on live data, a staged rollout and a prospective evaluation are the usual route into practice.
  • During research, the most useful preparation is a clinical owner, a written intended use, good records and an external test.

This article is part of the AI in Clinical Research Knowledge Base.