Human-AI Interaction and User Interface Design
How the presentation of AI results affects clinical decision-making. Alert fatigue, anchoring bias, automation complacency, and evidence-based principles for designing effective AI interfaces.
Most AI tools in medicine don't make decisions. They show something to a clinician, who makes the decision. So the thing that affects the patient is the combined performance of the model and the person, and that depends heavily on what is shown, when, and how.
It would be convenient if a good model plus a good clinician always added up to something better than either. The studies say that it often does, sometimes doesn't, and that the difference comes down to design choices that are usually made in a hurry at the end of the project.
The clinician plus the model is a third thing
In 2020 Philipp Tschandl and colleagues published a large study of clinicians diagnosing skin lesions with AI support. With good AI advice, accuracy improved, and the least experienced clinicians gained the most. They also tested what happened when the AI was deliberately made faulty. Accuracy fell, for experts as well as novices. And the form of the advice made a difference: showing the model's probabilities for each diagnosis helped, while some other formats did not.
So adding AI to a clinician can raise or lower their performance, depending on the model's quality on that case and on how the output is presented. We can't assume the clinician will filter out the model's mistakes. That assumption is behind the phrase "human in the loop," which is often offered as if it were a safety mechanism. It is one only if the human can tell when the model is wrong.
Ways it goes wrong
Automation bias
Automation bias is the tendency to go along with an automated suggestion, including when it is wrong. It has been studied in aviation for decades. Kate Goddard and colleagues reviewed the evidence in health care in 2012 and found it in a range of decision support systems. It comes in two forms. In errors of commission, the person follows wrong advice. In errors of omission, the person misses something because the system didn't flag it.
A 2023 study by Thomas Dratsch and colleagues tested this in mammography. Twenty-seven radiologists read mammograms along with a BI-RADS category said to come from an AI system, which for some cases had been made deliberately wrong. When the suggestion was correct, the least experienced readers got about 80% of cases right. When it was wrong, they got about 20% right. Very experienced radiologists were pulled along too, dropping from about 82% to about 46%.
Experience helps and doesn't make anyone immune. Neither does being told to be careful.
Anchoring
Seeing the model's answer before forming your own changes what you see. This is anchoring, and it is why the order of events makes a difference. A reader who looks at the image first and then checks the AI is doing a different task from one who sees the AI's marks from the start.
Alert fatigue
When a system interrupts often and is usually wrong or irrelevant, people stop reading the interruptions. The classic evidence is from drug safety alerts in prescribing systems. Heleen van der Sijs and colleagues reviewed the literature in 2006 and found that clinicians overrode between 49% and 96% of them. Once dismissing alerts is a reflex, the important one gets dismissed along with the others.
An AI tool that flags too much joins that pile. The alert rate at real prevalence is worth calculating before deployment (see 8.2 Clinical Workflow Analysis).
Complacency and lost skills
People who are assisted by a reliable system gradually check it less. There is also some early evidence that skills fade. In 2025 Krzysztof Budzyń and colleagues reported on endoscopists at centres in Poland that had introduced AI-assisted polyp detection. When those endoscopists later performed colonoscopies without the AI, their adenoma detection rate was about 22%, down from about 28% before they had started using it. It was an observational study and one result, so we wouldn't lean on it too hard, but it's a question worth asking of any tool people will come to depend on.
Misplaced faith in explanations
The usual proposed fix for all of this is to show the clinician why the model said what it said, on the theory that they'll then spot bad reasoning. The evidence for that is not encouraging. In a 2023 study, Sarah Jabbour and colleagues gave clinicians patient vignettes with AI predictions, some from a model that had been made systematically biased. The biased model lowered clinicians' diagnostic accuracy, and adding image-based explanations of the model's reasoning didn't rescue them.
As 1.6 What AI Cannot Do describes, heat maps can look plausible whatever the model is doing. An explanation can make a wrong answer more persuasive (see 7.4 Explainability and Interpretability in Clinical AI).
Disuse
The opposite failure gets less attention and is at least as common. Clinicians who don't trust a tool ignore it, and then whatever benefit it had is lost. That is the subject of 8.5 Clinician Trust, Adoption, and Change Management.
The design choices
When the output appears
There are three basic arrangements for an AI tool that assists a reader.
In a second-reader design, the clinician forms their own opinion first and then sees the AI's output. This protects against anchoring and keeps the clinician's skills in use. It takes more time.
In a concurrent design, the AI's output is visible from the start. It is faster, and it is the arrangement most exposed to automation bias.
In a triage design, the AI decides the order of cases, or which cases get a human look at all, and the clinician doesn't see its output on the individual case. Reader behaviour on each case is unaffected. The risk moves to the cases that were ranked low.
We don't think one of these is right in general. The choice should be made on purpose, with the trade written down, and the evaluation should use the same arrangement that will be deployed. A reader study done in second-reader mode says little about a product used concurrently (see 5.5 Reader Studies).
What is shown
A model can report a yes or no, a score, a category, a location, an outline, a measurement, or some combination. A few things are worth knowing when choosing.
A bare yes or no hides how sure the model is, and treats a borderline case the same as an obvious one.
A probability is only helpful if it is calibrated, meaning that cases scored 80% are positive about 80% of the time (see 4.4 Calibration). People also read numbers inconsistently. Grouping scores into a few named bands that each map to an action ("no flag," "review," "urgent") is often easier to use than a raw percentage.
Marks on the image are quick to check, which is a real advantage: the clinician can look at the spot and agree or not. They also draw the eye, and regions the model didn't mark get less attention.
Whatever the format, less is usually better. A display with a score, a heat map, three similar cases and a paragraph of text takes time to read, and the clinician has perhaps a few seconds to spare for it.
Say what the tool can't do
Users need to know the tool's limits at the point of use: which patients and image types it was validated on, what it doesn't look for, and how well it performed. A fracture detector that doesn't assess the spine should say so somewhere more visible than page 40 of the manual, or its silence on a spinal fracture will be read as reassurance.
Even better is a tool that declines to answer when the input is outside what it knows: the wrong body part, poor image quality, a patient outside the validated age range. "No result: image quality insufficient" is a useful output.
Make disagreeing easy
If accepting the AI's suggestion takes one click and rejecting it takes five and a typed justification, the interface is pushing people toward acceptance. Rejecting or editing a result should take no more effort than accepting it.
Recording those rejections is worth doing too. They are the cheapest source of information about where the model is failing in practice (see 8.6 Monitoring AI Performance in Production).
Interrupt sparingly
A result that waits in the worklist or the viewer costs the clinician nothing until they look at it. A pop-up, a page or a phone call costs attention immediately. Interruptions are best saved for findings that are both urgent and reasonably certain, and everything else can be passive.
Test the interface with people
The only way to know how a design behaves is to watch clinicians use it. This is called usability testing, and for medical devices it is a regulatory expectation as well as a good idea. The standard is IEC 62366, and the FDA has guidance on applying human factors engineering to devices (see Appendix C).
Early on, that can be as simple as sitting five clinicians in front of a prototype with realistic cases and asking them to think aloud. Before release, it means a more formal test with representative users doing the critical tasks.
Two things are worth building into those tests. Include cases where the AI is wrong, and see what the users do. A test in which the AI is always right tells you nothing about the situation you most need to understand. And measure the performance of the clinician with the tool, since that is what reaches the patient.
It also pays to have people on the team who design interfaces for a living. Researchers and engineers building a display for other clinicians will tend to build the one they themselves would like, which shows everything.
Key takeaways
- The clinical result depends on the clinician and the model together, and that combination can be better or worse than the clinician alone.
- Automation bias leads people to follow wrong AI advice. It affects experienced clinicians too, and a human in the loop is a safeguard only when the human can tell that the model is wrong.
- Frequent or low-value alerts get ignored, and the important ones go with them.
- Explanations such as heat maps have not been shown to protect clinicians from a wrong model, and can make wrong output more convincing.
- The timing of the output is a deliberate choice among second-reader, concurrent and triage designs, and the evaluation should match the deployment.
- Show a small amount of calibrated information, state what the tool doesn't cover, and let it decline to answer on unsuitable inputs.
- Make rejecting the AI as easy as accepting it, and log the rejections.
- Test the interface with real users, including on cases where the AI is wrong.
What to read next
- 8.5 Clinician Trust, Adoption, and Change Management: getting people to use the tool appropriately
- 8.2 Clinical Workflow Analysis: Where Does AI Fit?: where in the process the output belongs
- 5.5 Reader Studies: Measuring AI's Impact on Clinical Decision-Making: measuring the clinician and model together
- 7.4 Explainability and Interpretability in Clinical AI: what explanations can and can't tell a user
- 4.4 Calibration: Does a 90% Confidence Score Really Mean 90%?: making the displayed number mean something
This article is part of the AI in Clinical Research Knowledge Base.