Reader Studies: Measuring AI's Impact on Clinical Decision-Making
Designing and interpreting reader studies (with-AI vs. without-AI paradigms). Crossover designs, washout periods, reader selection, and measuring the effect on diagnostic accuracy and efficiency.
Most medical AI is sold as an assistant. The model marks something, scores something or drafts something, and a clinician makes the call. For a tool like that, standalone accuracy is only half of the picture. The other half is what the clinician does when the output is in front of them, and the way to measure that, short of a clinical trial, is a reader study.
In a reader study a group of clinicians interprets a set of cases under two conditions, with the AI and without it, and you compare their performance. It's a laboratory experiment on human decision-making, which makes it a different kind of study from anything in section 4, and the design details have a large effect on what the result means.
What a reader study can tell you
There are three quantities, and a good study reports all of them on the same cases.
- The model alone (standalone performance).
- The readers alone.
- The readers with the model.
The headline comparison is readers with against readers without. That's the claim an assistive product makes, and it's what the FDA generally wants to see for computer-assisted detection and diagnosis software.
The other comparisons are informative too. If readers with the model do worse than the model alone, the readers are overriding correct outputs. If readers with the model do no better than readers alone, despite a strong model, they are ignoring it, or it's finding the same cases they find. None of that can be seen from a standalone AUC.
The MRMC design
The standard design is called multi-reader multi-case, or MRMC. Several readers each read many cases, and in the usual "fully crossed" version every reader reads every case in both conditions.
The reason for crossing is statistical efficiency. Readers differ a great deal from one another, and cases differ in difficulty. Having each reader serve as their own control, on the same cases, removes most of that variation from the comparison.
The reason for having many of each is generalization. The claim you want to make is about radiologists in general reading patients in general, so both readers and cases are treated as random samples from larger populations. A study with three readers supports a conclusion about those three people, whatever the number of cases. This also means MRMC data needs its own analysis methods, which account for variation among readers, among cases, and the correlation created by everyone reading the same cases. The two classic approaches are named for their authors, Dorfman, Berbaum and Metz, and Obuchowski and Rockette, and the FDA distributes free software for MRMC analysis and sample size planning. This is a job for a statistician who has done it before.
As a rough guide, published reader studies supporting regulatory submissions tend to have somewhere between ten and twenty-odd readers and a few hundred cases. Adding readers usually buys more power than adding cases once you have a couple of hundred.
Order and memory
A reader who sees a case twice may remember it the second time. There are two common ways to handle this.
In a crossover design, readers interpret all the cases in one condition, wait, and then interpret them again in the other. Half the readers start with AI and half without, and cases are presented in a different random order in each session. The wait is called a washout period, and four weeks or more is typical. This design gives the cleanest independent estimates of the two conditions, and it costs calendar time.
In a sequential design, the reader interprets each case unaided, records an answer, is then shown the AI output and may revise. It takes a single session and needs no washout. It matches the workflow of a tool used as a second look. It doesn't match a tool whose output is visible from the start, and the unaided read tends to be a little more careful than usual, since the reader knows a machine is about to check their work.
Pick the one that mirrors how the product will be used. If the output will be on screen when the image opens, test it that way.
Choosing readers
The readers should be the people who will use the tool. If it's meant for general radiologists, a panel of subspecialty-trained academic experts tells you about a population that needs the tool least. If it's meant for emergency physicians reading their own films, test emergency physicians.
Include a range of experience, and record it, because the effect of AI assistance often differs between readers. The intuitive expectation is that juniors gain the most. That is sometimes what's found, though not reliably. In 2024 Feiyang Yu and colleagues reported a study of 140 radiologists reading chest radiographs with AI assistance, and found that the effect varied widely from one radiologist to the next and that experience and subspecialty were poor predictors of who was helped. So we'd suggest planning to report the per-reader results as well as the average.
Give the readers proper training on the tool before the study, including what it does and doesn't detect and how often it is wrong. Untrained readers produce an estimate of how the tool performs in its first confused hour of use.
Choosing cases
Reader study case sets are almost always enriched. At screening prevalence a set of 300 mammograms would contain one or two cancers, and nothing could be estimated. So half or a third of the cases are positive.
Enrichment is necessary, and it changes the readers' behaviour, since they quickly notice that disease is everywhere and adjust their threshold. Both conditions are affected, so the comparison between them survives better than the absolute numbers do. The absolute sensitivities and specificities from an enriched reader study shouldn't be quoted as estimates of clinical performance.
Within the enrichment, the mix of cases makes a big difference. A case set of obvious positives and clean normals leaves no room for the AI to help, or for it to mislead. Include subtle positives, and include negatives with benign findings that could be mistaken for the disease. Deliberately include cases the model gets wrong, in roughly the proportion it gets them wrong on unselected data, since a case set made entirely of the model's successes guarantees a flattering result. And the cases must be independent of the model's training data.
The reference standard for the cases needs to be stronger than the readers themselves: pathology, follow-up, or a separate expert panel with more information than the study readers get (see 5.7 Reference Standards and Ground Truth in Medical AI).
What to measure
Accuracy first. Depending on the task this is AUC from readers' confidence ratings, sensitivity and specificity from their binary decisions, or a localization measure when finding the lesion in the right place is the point. If readers must mark the location, a correct "abnormal" call that points at the wrong spot shouldn't count.
Reading time is the usual secondary endpoint, and for some products it's the main claim. It is hard to measure well in a lab, since readers aren't being interrupted by phone calls, and the first few dozen cases with a new tool are slower than the hundredth.
Confidence, and agreement between readers, are often collected too. AI assistance tends to make readers agree with each other more, which is good when the model is right.
When the model is wrong
The average effect of assistance hides the cases of most interest: the ones where the model's output was incorrect. Did readers who would have been right on their own change their answer to match it?
This is automation bias, and reader studies are the main way it has been measured. In a 2023 study by Thomas Dratsch and colleagues, 27 radiologists read mammograms with a suggested BI-RADS category said to come from an AI system. When the suggestion was deliberately wrong, accuracy fell at every level of experience, and it fell furthest for the least experienced readers.
So analyse the cases the model got wrong separately, in both directions: correct decisions that were overturned by a wrong output, and wrong decisions that were rescued by a right one. A tool with a good average effect and a large number of overturned correct decisions has a design problem in how its output is presented (see 8.4 Human-AI Interaction and User Interface Design).
What a reader study isn't
A reader study happens in a laboratory. David Gur and colleagues showed in 2008 that radiologists reading mammograms in a retrospective study behaved differently from the same radiologists in clinical practice, and they called this the laboratory effect. Readers in a study know that no patient is affected by their decision, that disease is common in the set, and that they are being watched. They usually have less clinical information than in practice, no priors, and no colleague to ask.
We already described, in 5.1, how mammography CAD looked helpful in early studies and failed to improve screening in practice. Reader studies were part of that early evidence.
None of this makes the design useless. It makes it a controlled experiment that isolates one question, whether the output changes decisions for the better under good conditions. A positive result is a reason to go on to prospective evaluation (see 5.3 Designing Prospective Clinical Validation Studies), and a negative one will probably save you the cost of a trial.
A planning checklist
Before the first reader sits down, we'd want the protocol to specify:
- The primary comparison and endpoint, and the analysis method.
- The number of readers and cases, with a power calculation based on a pilot or on published variance estimates.
- Who the readers are, how they were recruited, and their training on the tool.
- How cases were selected, the enrichment, and the proportion on which the model is wrong.
- The reference standard and who established it.
- Crossover or sequential, the washout period, and the randomization of order.
- What clinical information readers get, and the viewing conditions.
- The model version and how its output is displayed.
- The planned analyses by reader experience and of cases where the model erred.
Key takeaways
- A reader study compares clinicians' performance with and without the AI on the same cases. It's the standard evidence for an assistive tool.
- Report three things on the same cases: the model alone, readers alone, and readers with the model.
- The multi-reader multi-case design treats both readers and cases as random samples, and needs specialized analysis. Plan for ten or more readers and a few hundred cases.
- Use a crossover design with a washout of about four weeks or a sequential design, whichever matches how the tool will be used.
- Recruit readers who resemble the intended users, across a range of experience, and train them on the tool.
- Case sets are enriched, so absolute accuracy figures don't transfer to practice. Include subtle cases and cases the model gets wrong.
- Analyse the cases where the model erred, to measure automation bias.
- Laboratory reading differs from clinical reading. A reader study supports going on to prospective evaluation and doesn't replace it.
What to read next
- 5.3 Designing Prospective Clinical Validation Studies: testing the same question in real practice
- 5.7 Reference Standards and Ground Truth in Medical AI: the truth the readers are scored against
- 4.7 Comparing Models: When Is One Model Truly Better Than Another?: comparing AI with clinicians fairly
- 8.4 Human-AI Interaction and User Interface Design: how presentation changes what readers do
- 8.5 Clinician Trust, Adoption, and Change Management: over-trust and under-trust outside the lab
This article is part of the AI in Clinical Research Knowledge Base.