A retrospective study tells you how a model does on archived cases. A prospective study runs the model on new patients as they arrive, in the setting where it will be used, and that is a much harder test. The inputs aren't curated, the equipment is whatever is in the room, and in the fuller designs there are clinicians reacting to the output.
You know how to design a prospective study. What follows is mostly about the design decisions that change when the intervention is software.
Do you need one?
Not always, and since prospective studies are expensive it's worth asking the question properly.
A retrospective study is often enough when the model's output is a measurement or a prioritization that a clinician checks as a matter of course, when a wrong output is cheap and quickly spotted, and when the retrospective data really do look like the inputs the model will get. Plenty of imaging AI products have been cleared by the FDA on retrospective evidence.
You probably need prospective evidence when:
- The model acts without a clinician reviewing each output, as in autonomous screening.
- The benefit depends on something a retrospective study can't observe: whether an alert is acted on, whether the tool changes behaviour, whether time is saved.
- The inputs are acquired differently in real use. A model that guides image acquisition, or that depends on photographs taken by non-specialists, can't be tested on images an expert already took.
- The claim is about outcomes.
- The product is first of its kind and the regulator has no predicate to reason from (see 6.4 The De Novo Pathway). If a submission is the goal, a Q-Submission meeting is the place to find out what the FDA expects before you design the study.
Three kinds of prospective study
Silent mode
The model runs live on real patients, and nobody sees its output. Results are logged and compared later with the reference standard and with what the clinicians did.
This is sometimes called shadow mode, and it is the cheapest prospective design by a long way. It tests the plumbing (does the right series reach the model, how often does it fail, how long does it take) and it measures accuracy on truly unselected live data. Because care isn't affected, the ethics review is usually light. We'd suggest it as a default first step before anything interventional, since integration problems found in silent mode are much less costly than the same problems found during a trial.
Its limit is the obvious one: no clinician used the output, so it says nothing about effect.
Prospective diagnostic accuracy study
Patients are enrolled consecutively, the model's output is recorded, and every patient gets the reference standard. The model's output may or may not be shown to the treating team.
The pivotal trial of IDx-DR, the autonomous diabetic retinopathy system, is the standard example. Michael Abràmoff and colleagues enrolled 900 people with diabetes at ten primary care sites. Staff with no ophthalmic imaging experience took the retinal photographs after a short training, the system produced a result, and every participant was also photographed and graded to a research standard by a reading center. The sensitivity and specificity endpoints were set in advance, at 85% and 82.5%, and the system came in at about 87% and 91%, with a gradable result in 96% of participants. The FDA granted a De Novo authorization in 2018.
Notice what the design tested besides the algorithm: untrained operators, primary care lighting, real patients with small pupils and cataracts. Compare that with the Thai clinic deployment described in 1.6 What AI Cannot Do, where a fifth of images were rejected as ungradable. Those are the things a prospective accuracy study exists to find out.
Randomized trial
Patients, clinicians or sites are randomized to care with the tool or care without it, and you compare an endpoint between arms. This is the only design that measures what the tool causes.
The largest group of AI trials so far is in colonoscopy. In one of the first, published in 2019, Pu Wang and colleagues randomized about a thousand patients to colonoscopy with or without a real-time polyp detection system and found an adenoma detection rate of 29% with the system against 20% without. Dozens of trials of similar systems have followed.
Outside endoscopy, the EAGLE trial at Mayo Clinic randomized 120 primary care teams, covering more than 22,000 patients, to receive or not receive an AI result on routine ECGs flagging probable low ejection fraction. New diagnoses of low ejection fraction were more frequent in the intervention arm. And the MASAI trial, mentioned in 5.1, randomized about 80,000 women in a screening program.
Design decisions that differ for AI
What exactly is the intervention?
In a drug trial the intervention is the drug. In an AI trial it is the model, plus how its output is displayed, plus the training users were given, plus what the protocol says they should do with a positive result. Change any of these and you have a different intervention.
So the protocol needs to describe all of it, including the software version. The SPIRIT-AI extension for trial protocols and the CONSORT-AI extension for trial reports, both published in 2020, list the items to specify: the version, how inputs were acquired and selected, how the output was presented, how humans interacted with it, and how errors and failures were analysed (see 9.1 Reporting Standards).
Freeze the model
The model, threshold and interface should be locked for the duration. Software is easy to update, and a well-meaning engineer who fixes a bug halfway through has split your trial into two smaller trials of two different devices. If an update is unavoidable for safety reasons, record the date and plan a sensitivity analysis. Models that keep learning during a trial need that specified in the protocol from the start.
The unit of randomization
Randomizing individual patients is the most efficient, and it works when each use of the tool is self-contained. It often isn't. A radiologist who reads with AI support in the morning carries what they learned into the afternoon's control cases, and an alert system installed on a ward affects everyone on it. This contamination pushes the two arms together and understates the effect.
The usual answer is to randomize clusters: clinicians, teams, clinics or time blocks. EAGLE randomized care teams for this reason. Cluster designs need more participants for the same power, and the analysis has to account for the clustering, which your statistician will want to hear about early. Stepped-wedge designs, where every site eventually gets the tool in a random order, are popular with hospitals that want everybody to have it in the end, and are harder to analyse when practice is changing over time anyway.
Blinding
You usually can't blind the clinician to whether they have the tool. What you can do is blind the people who assess the outcome and the people who determine the reference standard, and choose endpoints that are objective. Where the endpoint involves judgment (was this polyp an adenoma?) the assessor should not know the arm. Pathologists reading the polyps in the colonoscopy trials are the model for this.
Knowing they are being observed with a new gadget also changes how clinicians behave, in both arms. A run-in period helps a little, and so does a long enough study that the novelty wears off.
The control arm
Usual care is the natural control, and it needs to be real usual care. If the control arm gets extra attention, double reading or protocol-driven care that isn't standard, the trial measures the tool against a comparator nobody uses. On the other hand, if the comparator is unusually weak (inexperienced readers, say), the tool looks better than it would in most hospitals.
Enrichment
When the condition is rare, an unselected cohort needs to be enormous to yield enough events. You can enrich by recruiting from a higher-prevalence setting, such as a referral clinic in place of a screening population. That is legitimate and common, and it has the same cost as in retrospective work: the spectrum of disease shifts toward the more obvious, and predictive values no longer apply to the intended population. State the enrichment and, if the claim is about screening, expect to be asked for unenriched evidence as well.
Endpoints and sample size
For an accuracy study the endpoints are typically sensitivity and specificity as co-primary, each tested against a goal fixed in advance, and the sample size is driven by the number of positive cases you need (see 4.5 Statistical Significance, Confidence Intervals, and Sample Size for AI Studies). For a trial, the endpoint should be the thing the tool is supposed to improve, as far along the causal chain as you can afford. 5.6 Defining Clinically Meaningful Endpoints is about that choice.
Either way, also plan to measure how the tool was used: how often the output was produced, how often it was looked at, and how often the clinician acted in line with it. A null trial with 20% uptake means something very different from a null trial with 95% uptake.
Safety
Software doesn't cause adverse events directly, so people sometimes leave this section thin. The harms are the consequences of wrong outputs: the missed case that was reassured, the false positive that led to a biopsy. Pre-specify how you'll capture them, and set up a way to review the model's errors as the study goes, since a systematic failure (a scanner software update that breaks the input, for instance) should be caught in weeks.
A middle stage worth knowing about
Between silent mode and a full trial there is room for a small, early, live evaluation: the tool in real use by a handful of clinicians, with attention on usability, workflow fit, the learning curve and safety. The DECIDE-AI guideline, published in 2022, was written for reporting this stage, which its authors compare to the early-phase studies of surgical innovation. If the tool is going to fail because its alerts arrive after the decision has been made, we'd much prefer to learn that from twenty clinicians than from a two-year trial.
Ethics and consent
Silent-mode studies generally qualify for a waiver of consent, since care is unchanged. Once the output influences care, the ethics board will want to know what happens when the model is wrong, who is responsible for the decision, and how patients are told. Cluster trials of tools embedded in routine care have sometimes been run with waived or modified consent, on the argument that individual consent is impracticable and risk is minimal. That is the board's call, and the conversation is easier when started early (see 7.6 Ethical Frameworks for AI in Healthcare).
Key takeaways
- Prospective evidence is needed when the model acts autonomously, when the benefit depends on how people respond to it, when real-use inputs differ from archived ones, or when the claim is about outcomes.
- Silent mode, where the model runs live and nobody sees the output, is a cheap first step that tests integration and live accuracy.
- A prospective diagnostic accuracy study tests the model with real operators and real acquisition conditions against pre-specified performance goals.
- A randomized trial is the only design that measures the effect of using the tool.
- The intervention is the model plus its interface, user training and response protocol. Specify all of it, freeze the software version, and follow SPIRIT-AI and CONSORT-AI.
- Contamination between arms is common with AI tools, so cluster randomization is often the better choice.
- Blind outcome assessors and reference standard readers even when users can't be blinded.
- Measure uptake and adherence alongside the endpoint, and plan how the model's errors will be captured as harms.
What to read next
- 5.5 Reader Studies: Measuring AI's Impact on Clinical Decision-Making: a controlled alternative when a trial is out of reach
- 5.6 Defining Clinically Meaningful Endpoints: choosing what the study measures
- 8.2 Clinical Workflow Analysis: Where Does AI Fit?: designing the intervention around the workflow
- 9.1 Reporting Standards: CONSORT-AI, SPIRIT-AI, TRIPOD+AI, and STARD: the checklists to design against
- 6.12 Post-Market Surveillance and Real-World Monitoring: prospective evidence after launch
This article is part of the AI in Clinical Research Knowledge Base.