Reference Standards and Ground Truth in Medical AI

The challenge of defining "truth" when expert clinicians disagree. Consensus panels, adjudication protocols, imperfect reference standards, and their impact on reported model performance.


Every accuracy figure in a medical AI paper is a comparison against something. Engineers call that something ground truth, a term borrowed from remote sensing, where you can walk out and check what is on the ground. Clinical researchers call it the reference standard, a more modest name that admits it is the best available method and may be wrong.

The modest name is the right one. In most medical AI studies the reference standard is expert opinion, and experts disagree. This article is about what that does to your results and how to build a reference standard for a validation study that will hold up. 2.3 Data Annotation and Labeling for Clinical AI covered the mechanics of labeling, agreement statistics and adjudication for training data. Here we're concerned with the test set, where the bar is higher.

Why the test set gets special treatment

A model can learn fairly well from imperfect labels, especially when the errors are random. A test set can't measure well with them. Errors in the reference standard change the measured performance directly, and they can change it in either direction.

So the two kinds of label usually come from different processes. Training labels can be single reads, or even labels mined from reports. Test labels should come from the strongest source you can afford, applied to every case, by people who haven't seen the model's output.

What an imperfect reference does to the numbers

Here's a worked example with invented numbers.

Take 1,000 cases, of which 200 truly have the disease. Suppose we have a perfect model: it flags exactly those 200. Our reference standard is an expert panel that is very good but human. It misses 10% of real disease, so it calls 180 of the 200 positive, and it wrongly calls 5% of the healthy cases positive, which is 40 of the 800.

Now score the perfect model against the panel. The panel says 220 cases are positive (180 real, 40 not). The model flagged the 180 real ones and none of the 40, so its measured sensitivity is 180 out of 220, or 82%. The panel says 780 are negative, and the model flagged the 20 of those that were real disease the panel missed, so measured specificity is 760 out of 780, about 97%.

A perfect model has been measured at 82% sensitivity, and every "error" it made was the panel's.

That is the case where the reference standard's errors are unrelated to the model's, and it pushes measured performance down. The other case is less comfortable. Suppose the model was trained on labels from these same experts and has learned their habits, including the kinds of lesion they overlook. Then it agrees with the panel on the panel's mistakes, and scores close to 100% while missing 10% of disease. Errors that the model and the reference standard share make performance look better than it is.

So an imperfect reference standard can bias the result down or up, depending on whether its errors are independent of the model's. Since models are usually trained on labels from the same kind of expert who provides the reference standard, the optimistic direction deserves more worry than it usually gets.

The hierarchy, again

2.3 ranked label sources for trustworthiness, and the same ranking applies here with more at stake. In descending order: an independent definitive test, clinical outcome over adequate follow-up, a panel of experts with more information than the model had, a panel of experts looking at the same images, a single expert.

The underlying principle is independence. The best reference standard uses information the model didn't have: tissue, a different imaging modality, time. A reference standard made by experts looking at the same image the model saw can only tell you whether the model agrees with experts.

Even the top of the hierarchy has some wobble in it. Pathology is treated as truth for most imaging studies, and pathologists disagree with each other too. In a well-known 2015 study led by Joann Elmore, pathologists interpreting breast biopsy slides agreed with an expert consensus diagnosis about 75% of the time overall, and under half the time for atypia. And any reference standard that depends on biopsy is only available for patients someone chose to biopsy, which brings in the verification problems described in 5.2 Designing Retrospective Validation Studies.

Building an expert panel

When opinion is the only available reference, the aim is to make it as good and as independent as possible. These are the design choices.

Use several readers. Three is the usual minimum, since a single reader's idiosyncrasies become the standard and two can only tie. Readers should read independently first, without seeing each other's answers.

Use readers who are at least as expert as the tool's intended users, and preferably more. In a reader study the reference panel must be separate from the study readers and should outrank them in subspecialty experience.

Give the panel more than the model gets. This is the cheapest way to strengthen an opinion-based standard. The model sees one chest radiograph. The panel can have the prior films, the follow-up CT, the clinical course and the discharge diagnosis. They are then answering "what did this patient have?" and not "what do I see in this image?", which is a different and better question for a reference standard.

Keep the panel blind to the model's output, and also to the original clinical report if the model was trained on such reports. A panel that sees the AI's marks will tend to agree with them. This is incorporation bias, and it inflates accuracy.

Resolve disagreements by a method fixed in advance: majority vote, discussion to consensus, or referral to a senior adjudicator. 2.3 describes the trade-offs. For a test set we'd lean toward discussion or adjudication for the disagreed cases, since they are a minority of cases and carry a large share of the uncertainty.

Give the panel the same written definitions the annotators had, and allow an "indeterminate" answer.

Composite and differential standards

Often no single reference test can be applied to everyone. A composite reference standard combines several sources by a rule set in advance: positive if pathology is positive, or if follow-up imaging within twelve months shows progression; negative if pathology is negative or two years of follow-up are clear.

That's a reasonable design, and the important thing is that the rule is written before the results are known and doesn't include the model's output. It does mean different patients were verified by different methods with different error rates, which the paper should state. Where the choice of reference test depended on the result of the test being evaluated, or of something closely related to it, the estimates are biased, and there are statistical corrections for partial verification that your statistician will know.

The cases nobody can call

Some cases remain ambiguous after every expert has looked. There are three things you can do with them, and one of them is a bad idea.

The bad idea is to drop them. It's very common, usually described as "cases without consensus were excluded." Those are the hardest cases in the dataset, so removing them makes any model look better and makes the test set less like the clinic. If you exclude them, report how many there were, and give a sensitivity analysis with them included.

The second option is to force a label by adjudication and accept that it is uncertain. That's fine as long as the paper reports how many cases needed it.

The third is to keep the uncertainty. Report results separately for the unanimous cases and the disputed ones. Or score the model against each expert in turn and compare that with how the experts score against each other, which shows where the model sits in the range of human readers. For segmentation, comparing the model's Dice against each reader with the readers' Dice against one another does the same job.

How good can the number get?

Agreement among the experts sets a limit on what can be measured, as we said in 2.3. If panel members agree with the panel's own consensus 90% of the time, then a model at 90% is at the limit of what this reference standard can resolve. A model reported at 97% against a single reader, where readers agree with each other 85% of the time, has most likely learned that reader.

So a validation paper should always report how the reference standard itself performed: inter-reader agreement before adjudication, the number of cases adjudicated, and how many stayed indeterminate. It tells the reader how much to trust every other number in the paper.

There are statistical methods for estimating accuracy without a perfect reference, such as latent class analysis, which treats true disease status as an unobserved variable and estimates it from several imperfect tests together. They depend on assumptions, mainly about the independence of the tests' errors, that are hard to check. They can be useful as a sensitivity analysis. We wouldn't rely on them for a primary result.

When the label is a judgment call by nature

For some tasks disagreement comes from the definitions themselves. Grading scales with adjacent categories, tumor margins that fade into edema, "suspicious" versus "probably benign": two competent experts can look at the same thing and reasonably differ.

In these cases it helps to move the reference standard toward something harder where possible. A model for "suspicious nodule" can be validated against eventual malignancy. A model for retinopathy grade can be validated on the clinically actionable split between referable and non-referable, where agreement is much higher than on the exact grade. A segmentation can be validated on the derived volume or the treatment decision it feeds (see 5.6 Defining Clinically Meaningful Endpoints).

What regulators and reviewers ask

Expect questions about who established the reference standard and their qualifications, how many readers there were, how disagreements were resolved, what information they had, and whether they were blind to the device output. The FDA's public summaries for cleared AI devices routinely describe the "truthing" process, and the Good Machine Learning Practice principles call for reference datasets based on the best available methods (see 6.7 Good Machine Learning Practice). A written truthing protocol, produced before the reads, answers nearly all of it and is worth the afternoon it takes.

Key takeaways

  • The reference standard is the best available method for establishing the condition, and it has an error rate. In most AI studies it is expert opinion.
  • Reference errors that are independent of the model's errors make the model look worse. Errors the model shares with the reference, which is likely when both come from the same kind of expert, make it look better.
  • The strongest reference standards use information the model didn't have: pathology, another modality, follow-up.
  • For an expert panel, use three or more independent readers who are at least as expert as the intended users, give them more clinical information than the model gets, keep them blind to the model, and fix the adjudication method in advance.
  • Define composite reference standards before seeing results, and report which cases were verified by which method.
  • Excluding cases without consensus removes the hardest cases and inflates performance. Report them and analyse with and without.
  • Report the reference standard's own reliability. Agreement among the experts is the limit of what can be measured.